Skip to content
No.199 击穿电压测试仪
JIS C2110, (C3003), C3216-5, IEC-60851-5
This tester is for evaluating the dielectric strength (volt endurance/ volt destruction) of solid electric insulating material.
By exchanging the electrodes, the tester can also be used for board, sheet, textile, film, tape, tube, varnish, and paint.
Recent applications also include insulating film in lithium-ion batteries.
After setting the specimen between the two electrodes, voltage is applied mid-air or in-bath until the specimen insulation destructs.
The results are evaluated either by the insulation strength or destruction voltage level.
The tester is capable of both the short-time test and step-by-step test.
The tester can also be used for enamel coatings using the twistpair method by exchanging the electrodes.
Specification
Specifications are subject to change without notice.
Capacity
0.5 to 10 kVA
Maximum Output Voltage
AC 5 to 80 kV
Specimen
L500 mm (Twisting L 120mm), φ0.05 〜 φ3.2 mm
Test Method
Oil Bath or Air Bath
Standard Equipments
A-meter, V-meter, Volt-slider, Programmed Voltage Raising Device, Breakdown Current Regulator
Option
Timer, Mini Printer, Software
Safety Device
Safety Cover with Interlock Attached, Earth-Leakage Breaker, Upper Voltage Limit, Emergency Stop SW,
Warning Indicator & Buzzer
Electrode
Selected by Specimen
Power Source
AC 200 V, 1-Phase, 15 to 50 A, 50/60 Hz
Dimensions/ Weight (Approx.)
Differ by Specifications.