No.148-HD Specification
Specifications are subject to change without notice. | |
Stations | 3, 4 or 6 Stations |
Temperature Range | Max. 300°C (Oil Bath) |
Heating Speed | 120 ± 10°C/hr, 50 ± 5°C/hr, VICAT: 120 ± 10℃/hr, 50 ± 5°C/hr |
Bending Stress | JIS・ISO: 1.80MPa (Method A), 0.45MPa (Method B), 8.00Mpa (Method C) ASTM: 1.82Mpa, 0.455Mpa |
Weight Load | DTUL: Choose 2 weights from initial load 76.5gf to max. 3,210gf Options VICAT: 10 ± 0.2N, 50 ± 1N BP: 19.6 ± 20N |
Displacement Measurement | Differential Transformer: Scale 1/1,000mm, Stroke 4mm |
Pressure Foot | DTUL: R3.0 ± 0.2mm Options VICAT: 1.000 ± 0.015mm² BP: Ball Diameter 5mm |
Support Length | 64 ± 1mm, 100 ± 1mm Option 101.6 ± 0.5mm |
Refrigerating Device | Water Circulating System Option Built-In Refrigerator *For low test start temperature (e.g., 20 to 23°C) |
Oil Circulation Device | Propeller Type For 3 and 4 Stations: 2 Propellers For 6 Stations: 3 Propellers |
Software | Windows 10 compatible(Included) |
Accessories | Pressure Foot Adjustment, Specimen Holder |
Options | Silicon Oil, Safety Cover, Simultaneous Loading Device, Built-In Printer, Oil Level Sensor, Glass Sample for Calibration (Fused Quartz), Safety Door Lock, Signal Tower Light |
Power Source | Without refrigerator For 3 and 4 Stations: Single-Phase, AC200/220V 20A 50/60Hz For 6 Stations: Single-Phase, AC200/220V 30A 50/60Hz With refrigerator For 3 and 4 Stations: Three-Phase, AC200/220V 20A 50/60Hz For 6 Stations: Three-Phase, AC200/220V 30A 50/60Hz |
Dimensions & Weight (Approx.) | Floor Model W1,100 × D700 × H1,500mm 180 to 300 kg Tabletop Model W1,100 × D700 × H1,000mm 155 to 185 kg |